0

Multi-Laser / Multi-Detector Imaging with Amytracker

Amytracker are optotracers with structure-dependent photo-physical properties. All Amytracker variants are designed to bind to the Congo red binding pocket on the amyloid fibril and require a theoretical minimum of eight in-register parallel-β-strands for binding. Therefore, Amytracker reliably labels amyloids derived from a variety of amyloidogenic proteins or peptides from different species. Due to their structure-dependent photo physical properties, the Amytracker variants are only fluorescent when binding to a target and different targets can produce a difference in the molecules fluorescence spectrum. To investigate different targets, we recommend to perform imaging by exciting the sample with different wavelengths collecting fluorescence intensity in multiple emission ranges (see the table below for reference). Excitation- and emission spectra for all Amytracker variants can be accessed here.

Table: Multi-laser / Multi-detector imaging protocol performed on a Zeiss LSM800 CLSM.
Channel Excitation Emission range Amytracker variant
CH1 405 nm 400-490 nm Amytracker 480 Amytracker 680
CH2 405 nm 490-600 nm Amytracker 480 Amytracker 680
CH3 405 nm 600-660 nm Amytracker 480 Amytracker 680
CH4 405 nm 660-700 nm Amytracker 480 Amytracker 680
CH5 488 nm 500-580 nm Amytracker 520 Amytracker 540 Amytracker 630
CH6 488 nm 580-650 nm Amytracker 520 Amytracker 540 Amytracker 630
CH7 488 nm 650-700 nm Amytracker 520 Amytracker 540 Amytracker 630
CH8 561 nm 600-650 nm Amytracker 630 Amytracker 680
CH9 561 nm 650-700 nm Amytracker 630 Amytracker 680
CH10 Brightfield

Older Post